Background: Cardiac implantable electronic device (CIED) infections necessitate extraction and subsequent pacing interventions. Conventional methods after removing the infected CIED system involve temporary or semi-permanent pacing followed by delayed permanent pacemaker (PPM) implantation. Leadless pacemakers (LPs) may offer an alternative, allowing immediate PPM implantation without increasing infection risks.
Objective: To evaluate the safety and cost-effectiveness of LP implantation during the same procedure of CIED extraction, compared to conventional two-stage approaches.
Methods: Pacemaker-dependent patients with systemic or pocket infection undergoing device extraction and LP implantation during the same procedure at Sheba Medical Center, Israel, were compared to a historical group of patients undergoing a semi-permanent (SP) pacemaker implantation during the procedure, followed by a permanent pacemaker implantation.
Results: The cohort included 87 patients, 45 undergoing LP implantation and 42 SP implantation during the extraction procedure. The LP group demonstrated shorter intensive care unit days stay (1±3 days vs. 7±12 days, p<0.001) and overall hospital days (11±24 days vs. 17±17 days, p<0.001). Rates of infection relapse and one-year mortality were comparable between groups. Economic analysis revealed comparable total costs, despite the higher initial expense of LPs.
Conclusions: LP implantation during CIED extraction offers significant clinical and logistical advantages, including reduced hospital stays and streamlined treatment, with comparable safety and cost-effectiveness to conventional approaches.