Introduction:
Blood pressure (BP) measurement is a routine medical procedure, essential for monitoring both hypertension and hypotension, which are linked to high morbidity and mortality. Accurate and convenient BP monitoring remains a challenge, particularly relying on cuff-based devices or invasive measurements in postoperative and intensive care settings. This study aimed to assess the accuracy of BP readings obtained from a novel, cuffless, non-invasive photoplethysmography (PPG)-based chest patch monitor (Biobeat) compared to invasive arterial pressure (IAP) monitoring in cardiac surgery patients.
Methods:
This prospective, single-center study enrolled adults who had undergone cardiac surgery. A PPG-based chest patch was placed on the patients’ chest post-surgery, and the data were compared with simultaneous IAP measurements. Accuracy and agreement between the two methods were evaluated using Bland-Altman plots and Pearson`s correlations.
Results:
Ninety-six patients participated, with BP and heart rate (HR) data available for 88 and 92 patients, respectively. The average patient age was 63.2 years (±12.2), with 33% women. The average monitoring time was 25.6 hours (±17.2). A total of 78,659 systolic BP (SBP) readings, 78,818 diastolic BP (DBP) readings, and 92,544 HR readings were analysed. Excellent correlations were found for SBP (r=0.959), DBP (r=0.973), mean arterial pressure (MAP) (r=0.966), and HR (r=0.962). Bland-Altman analysis showed biases of 0.1 ± 4.8 mmHg for SBP, 0.4 ± 2.1 mmHg for DBP, 0.26 ± 2.6 mmHg for MAP, and 0.15 ± 3.6 beats per minute for HR. 86% of SBP measurements and all DBP readings were within 10 mmHg of the reference values. BP and HR changes measured by IAP were mirrored by the cuffless device.
Conclusion:
The cuffless device demonstrated high accuracy and agreement for BP and HR measurements compared to IAP monitoring in post-cardiac surgery patients. Further studies are required to confirm these findings across various clinical settings and populations using standardized validation protocols for cuffless devices